Jak do poprawnie Zrozumieć Metody Używanie pH Metry oraz Konserwacja Elektrody pH metry
① Napięcie i częstotliwość moc zasilanie mus przestrzegać z dane wskazane on instrument tabliczka, i it must be well grounded, otherwise the wskaźnik may be unstable during measurement
② The instrument is equipped with glass electrodes and calomel electrodes. Clamp the bakelite cap of the glass electrode onto the small clip of the electrode clamp. clamp the metal cap of the calomel electrode onto the large clamp of the electrode clamp. The height of the two elektrody can be adjusted using the support screws on the electrode clamp.
③ Szkło elektrody musi być moczone w destylowane woda dla co najmniej 24 godziny przed początkowe użycie. To powinno również być nasączone w destylowane woda kiedy nie w użycie.
Sekunda, kalibracja
② Włącz włącz zasilanie i wskaźnik światło wola światło up. rozgrzewka up for 30 minut.
③ Wziąć a mała zlewka zawierająca destylowana woda i delikatnie usuń nadmiar woda krople z szklana elektroda z filtr papier. Wlać a wybrane standardowe bufor roztwór z znane pH w a mała zlewka. Zanurz elektroda. Zwróć uwagę zanurz zanurz mała kulka at the koniec z szkło elektroda and kapilary pory z kalmel elektroda w roztwór. delikatnie wstrząsnąć mała zlewka do zrobić roztwór w kontakt z elektrodą mundurem.
④ According to the pH of the standard buffer, turn the range switch to positions 0-7 or 7-14.
⑤ Adjust the temperature control knob so that the temperature indicated by the : knob is the room temperature.
⑥ Adjust the zero point so that the pointer points at pH7.
⑦ Delikatnie naciśnij lub nieznacznie obróć the reading switch to make it stick. do the positioning knob so that the wskaźnik is exactly at the pH value of the standard buffer. Release the reading switch and repeat the operation until the value stabilizes.
Wreszcie, pomiar
① Zassać up nadmiar woda krople na elektroda lub płukać dwa razy z testowany roztwór, then immerse the elektroda in the tested solution and delikatnie rotate or shake the small beaker to make the the solution evenly contact the electrode.
② The temperature of the tested solution should be the same as that of the the standard buffer solution.
③ Adjust the zero position, press the reading switch, and the value indicated by the pointer is the pH of the liquid to be measured. If the pointer reading exceeds the scale when measuring with with the range of pH 0-7, the range switch should be placed at pH 7-14 before measurement.
④ After the measurement is completed and the reading switch is released, the pointer must point to pH7, otherwise readjust.
1. When the composite electrode is not used, it can be fully immersed in 3M Potassium chloride solution. Do not use washing
1. When the composite electrode is not used, it can be fully immersed in 3M Potassium chloride solution. Do not soak with detergent or other absorbent reagents.
2. Przed użycie, check the bulb on the front end of the glass electrode. Under normal conditions, the electrode should be transparent and free from cracks; The bulb should be filled with solution and no bubbles should exist.
Kiedy pomiar roztwory z wysokie stężenia, próba skrócenie pomiar czas jak dużo jak to możliwe oraz ostrożnie czyste ich po użyciu
4. Po czyszczeniu elektroda, do nie używanie filtr papier do wycieranie szkło folia. Zamiast tego, użycie filtr papier do absorbowanie i suche to uniknięcie uszkodzenie szkło folia, zapobieganie krzyżowe zanieczyszczenie, i wpływanie pomiar dokładność.
5. W trakcie pomiar, zwróć uwagę srebro srebro chlorek wewnętrzny odwołanie elektroda z elektroda powinien być zanurzony w chlorek bufor roztwór w żarówka do uniknąć the digital jump phenomenon in the electrometer display part. When useing, be careful to delikatnie shake the electrode a few times.
6. Electrodes should not be used in strong acids, alkalis, or other corrosive solutions.
7. It is strictly prohibited to use it in dehydrating media such as anhydrous ethanol, Potassium dichromate, etc.